Measuring biomass using optical backscattering
A brief introduction to biomass monitoring and backscattering measurements
During fermentation and cultivation of cells, yeasts and algae,<br>monitoring changes in biomass density is very important as it provides information about the growth and<br>overall health of the cell cultures.
Monitoring biomass can be used to assist controlling the process, optimising parameters for growth<br>and analysing and validating cultivation strategies.
Biomass monitoring produces a growth curve, i.e. how much has biomass concentration changed as function of time.<br>There are multiple methods for measuring biomass concentration, both optical and mechanical,<br>optical methods being far more prevalent. Optical methods are also readily suited for real-time measurements.
Growth curves
There are usually four sections or phases in a growth curve:
Inoculation: Cells ( bacteria, yeast etc. ) are introduced to the reactor
Lag phase: Cells adapt to their new environment before starting to multiply
Growth phase: Number of cells increases rapidly ( also called log phase )
Steady phase: Growth is limited by lack of resources
These curves are used to model system behaviour and to estimate parameters such as specific growth rate (µ) and doubling time (td).
Measurement methods
Three most common methods for optically measuring biomass density are:
Standard manual sampling
Sample is taken from the bioreactor and optical density is measured using a desktop spectrophotometer (transmission measurement)
Is quite labour intensive<br>--><br>Sampling consumes a small amount of growth medium and cells (an issue if cultivation volume is small)
Frequent sampling poses an increased contamination risk
Real-time transmission measurement
Rod-like sensor (optical measurement done at the tip of the rod) is installed into reactor vessel
Suitable only for larger reactors and vessels
Backscattering measurement
Uses light that is reflected (or more precisely, scattered back) from the cells
Can be retrofitted to existing setups<br>--><br>No risk of contamination as the sensor is not in contact with the growth medium
Requires calibration to provide OD600 readings comparable with manual sampling method
Backscattering measurement
In a backscattering measurement, a beam of light is emitted into the liquid containing cells. As light interacts with the cells, a small amount is scattered back to the sensor.<br>Intensity of this backscattered light is related to the number of scatterers in the beam ie. density of cells in the media.
Simpler geometry than with transmission where measured target must be between emitter and receiver
Receiver and emitter are next to each other and facing same direction
Sensor can be installed to side of a transparent vessel or to a window
Fast measurements
Multiple measurement per second
High time resolution can be used to monitor system dynamics
More frequent sampling is useful for data analysis
Dynamic range is very large, same sensor can be used to measure from near-transparent to high optical densities
Linearity of the sensor and rejection of ambient light is crucial
Unlike in transmission measurements where light is attenuated logarithmically , backscattering signal almost linear as optical density increases
To minimise unwanted effects caused by ambient light, a pulsed or chopped measurement is commonly used:
Signal is sampled at multiple points during a measurement pulse to remove ambient light contributions
Frequency of pulses is high enough so that even fast changes in prevalent lighting conditions are cancelled out
Statistical methods or time-domain filters are then applied to the signal, providing a more stable result
Contributing factors
Measured signal consists of reflections from the cells and environment:
Front surface (window / vessel surface), depends on the haziness of the window or surface
Liquid surface, if measurement beam reaches the surface of the medium
Back (or opposite) surface of the reactor vessel
Vessel floor, if visible to sensor
Other structures (stirrers, tubing etc.)
Larger vessel and higher optical densities reduce the effects of static structures, as light reaching and reflecting off of them spreads out more and is absorbed by the medium.
Note: Small vessel and very transparent liquid (i.e. optically thin scenario) is more challenging for backscattering measurements.
-->
Signal components
Imeas = Measured signal (i.e. light reflected back to sensor)
Iemit( A, T ) = Intensity of light emitted from sensor head
A = Target amplitude of light
T = Temperature at sensor head, higher temperatures yield less intensity
Gwindow = Window or vessel surface transparency
Higher transparency increases relative signal from within the...